Title: |
|
Philips HeartStart MRx Monitor/Defibrillators may fail to deliver therapy without alerting the user to a fault in the event of internal damage.
|
Broadcast content: |
|
The Philips HeartStart MRx Monitor/Defibrillator MRx may fail to identify a fault and alert the user in the event of internal damage suffered during a drop or due to severe mechanical shock.
|
Additional information: |
|
This alert is not relevant to Primary Care.
|
Action underway deadline: |
|
01-Jul-2020
|
Action complete deadline: |
|
15-Jul-2020
|