Title: |
|
Smiths Medical CADD® Non Flow-Stop Medication Cassette Reservoirs – recall of specific lots due to risk of under delivery of medication
|
Broadcast content: |
|
Smiths Medical Non Flow-Stop CADD® Medication Cassette Reservoirs may have been manufactured with an incorrect pressure plate which could cause partial or total occlusion of the tubing with no alarm. Please use the link below to access a spreadsheet for the list of affected lots:
https://www.gov.uk/drug-device-alerts/smiths-medical-cadd-non-flow-stop-medication-cassette-reservoirs-recall-of-specific-lots-due-to-risk-of-under-delivery-of-medication?cachebust=1529499546
|
Additional information: |
|
NHS Regional offices: do not cascade to primary care.
|
Action underway deadline: |
|
04-Jul-2018
|
Action complete deadline: |
|
18-Jul-2018
|