Action category: Action

Title: NIPPY ventilator range (all models) – update to instructions for use

Broadcast content: Manufactured by Breas Medical Limited – maintenance schedule now includes changing the internal memory/alarm battery every 3 years.

Additional information: NHS England Regional offices:
Do not cascade to primary care


Alert reference: MDA/2020/001

Action underway deadline: 22-Jan-2020

Action complete deadline: 26-Feb-2020