Title: |
|
IntelliVue MX40 patient-worn monitors – increased power consumption and no visual or audible alarms when batteries are low
|
Broadcast content: |
|
Manufactured by Philips – devices may lose power earlier than expected and users may not realise the loss of monitoring due to no alarm, which could contribute to a delay in emergency treatment.
|
Additional information: |
|
NHS England Regional offices: Do not cascade to primary care
|
Action underway deadline: |
|
23-Jul-2019
|
Action complete deadline: |
|
13-Aug-2019
|