Central Alerting System
View Alert


Originator: MHRA Medical Device Alerts

Issue date: 04-Mar-2019 11:00:07

This alert has been issued to:
  • Care Trusts
  • Mental Health Trusts
  • Specialists Trusts
  • NHS Regional Offices
  • Learning Disabilities Trusts
  • Mental Health & Social Care Trusts
  • Ambulance Trusts
  • Mental Health & Learning Disabilities Trusts
  • Acute Trusts
  • Community Trusts

  • Other contacts
  • Independent Healthcare Providers (registered with CAS)
  • Clinical Commissioning Groups
  • Ofsted recipients
  • Social Care Providers (registered with CAS)
  • Special Health Authorities

Action category: Action

Title: All T34 ambulatory syringe pumps need a sponge pad fitted to the battery compartment to prevent battery connection issues

Broadcast content: Manufactured by Caesarea Medical Electronics (CME) Ltd, a BD company – instructions provided to reduce the risk of delay to therapy and loss of infusion if the battery loses connection.

Additional information: NHS England regional teams, please cascade this alert to
• General practitioners (for information only)
• General practice managers
• General practice nurses


Alert reference: MDA/2019/013

Action underway deadline: 01-Apr-2019

Action complete deadline: 28-May-2019

Attachments:

Contact our helpdesk

Email: safetyalerts@mhra.gov.uk

Medicines and Healthcare products Regulatory Agency